User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: FLUKE 8810A AC/OHMS VERIFICATION DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 40 NUMBER OF LINES: 103 CONFIGURATION: Fluke 5200A/5215A Fluke 5450A ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD {AC VOLTAGE VERIFICATION} 1.002 OPBR Does this UUT have an AC VOLTS option, and if so ... 1.002 OPBR do you want to record data on it? 1.003 JMPF 23.004 1.004 OPBR Does this UUT have the AC AVERAGING (08) Option? 1.005 JMPF 14.004 1.006 DISP Connect the UUT LO and INPUT terminals to the 1.006 DISP ALTERNATING VOLTAGE CALIBRATOR OUTPUT terminals. 1.006 DISP Set the UUT to the ACV function. 1.007 5200 2 1V .0011U 400H 2W 2.001 5200 2 1V .0106U 100KH 2W 3.001 5200 20 10V .011U 400H 2W 4.001 5200 20 10V .106U 100KH 2W 5.001 5200 200 50V .06U 400H 2W 6.001 5200 200 50V .56U 100KH 2W 7.001 5200 750 100V .35U 400H 2W 8.001 5200 750 100V .65U 20KH 2W 9.001 ASK- R 9.002 DISP Set the UUT to AUTO RANGE. 9.003 5200 A 100mV .00045U 10KH 2W 10.001 5200 A 1V .0027U 10KH 2W 11.001 5200 A 10V .027U 10KH 2W 12.001 5200 A 100V .27U 10KH 2W 13.001 5200 * S 13.002 DISP Move the test leads from the ALTERNATING VOLTAGE 13.002 DISP CALIBRATOR to the PRECISION POWER AMPLIFIER OUTPUT. 13.003 5200 A 750V 2.27U 10KH 2W 14.001 5200 * S 14.002 ASK+ R 14.003 JMP 23.002 14.004 DISP This UUT has the TRUE-RMS AC VOLTS (09) option. 14.005 DISP Connect the UUT LO and INPUT terminals to the 14.005 DISP ALTERNATING VOLTAGE CALIBRATOR OUTPUT terminals. 14.005 DISP Set the UUT to the ACV function. 14.006 5200 2 .1V .0012U 500H 2W 15.001 5200 2 1V .003U 500H 2W 16.001 5200 2 1V .0055U 50KH 2W 17.001 5200 20 10V .03U 500H 2W 18.001 5200 20 10V .055U 50KH 2W 19.001 5200 200 100V .3U 500H 2W 20.001 5200 200 100V .55U 50KH 2W 21.001 5200 * S 21.002 DISP Move the test leads from the ALTERNATING VOLTAGE 21.002 DISP CALIBRATOR to the PRECISION POWER AMPLIFIER OUTPUT. 21.003 5200 750 500V 2U 500H 2W 22.001 5200 750 500V 3.5U 50KH 2W 23.001 5200 * S 23.002 HEAD 23.003 DISP Disconnect the UUT from the PRECISION POWER AMPLIFIER. 23.004 HEAD {RESISTANCE VERIFICATION} 23.005 OPBR Does this UUT have the OHMS (07) option, and if so ... 23.005 OPBR do you want to record data on it? 23.006 JMPF 40.003 23.007 DISP Connect the UUT to the RESISTANCE CALIBRATOR -- 23.007 DISP ** FOUR-WIRE MODE ** 23.007 DISP Set the UUT to the KOHM function. 23.008 DISP The 20M LED should be on for the next reading. 23.009 5450 20 10MZ .0203U 4W 24.001 EVAL Was the 20M LED on for the last test? 25.001 5450 2000 1000KZ .53U 4W 26.001 5450 200 100KZ .013U 4W 27.001 5450 20 10KZ .0013U 4W 28.001 5450 2 1KZ .00013U 4W 29.001 DISP The 200 OHM LED should be on for the next FOUR readings. 29.002 5450 200 100Z .03U 4W 30.001 5450 200 10Z .012U 4W 31.001 ASK- R 31.002 DISP Set the UUT to AUTO RANGE. 31.003 5450 A 10Z .012U 4W 32.001 5450 A 100Z .03U 4W 33.001 EVAL Was the 200 OHM LED on during the last FOUR readings? 34.001 5450 A 1KZ .00013U 4W 35.001 5450 A 10KZ .0013U 4W 36.001 5450 A 100KZ .013U 4W 37.001 5450 A 1000KZ .53U 4W 38.001 DISP The 20M LED should be on for the next test. 38.002 5450 A 10MZ .0203U 4W 39.001 EVAL Was the 20M LED on during the last test? 40.001 5450 * S 40.002 ASK+ R 40.003 HEAD 40.004 DISP Disconnect the UUT from the RESISTANCE CALIBRATOR. END